Caring for Your Test Strips
● Storage conditions: 2°C to 30°C (35.6°F to 86°F), 10% - 85%
relative humidity. Do
NOT
freeze.
● Store your test strips in their original vial only. Do not transfer to
another container.
● Store test strip packages in a cool dry place. Keep away from
direct sunlight and heat.
● After removing a test strip from the vial, immediately close the vial
cap tightly.
● Touch the test strip with clean and dry hands.
● Use each test strip immediately after removing it from the vial.
● Write the opening date on the vial label when you first opened
it. Discard remaining test strips after 3 months.
● Do not use test strips beyond the expiry date. This may cause
inaccurate results.
● Do not bend, cut, or alter a test strip in any way.
● Keep the strip vial away from children since the cap and the test
strip may be a choking hazard. If swallowed, promptly see a doctor
for help.
For further information, please refer to the test strip package insert.