22
NOTICE:
To avoid contaminating the control solution, do not directly apply
control solution onto a strip.
4. Read and compare the result
After counting down to 0, the control
solution test result will appear on the
display. Compare this result with the
range printed on the test strip vial and
it should fall within this range. If not,
please read the instructions again and
repeat the control solution test.
(118 mg/dL = 6.5 mmol/L; 118−160
mg/dL = 6.5−8.8 mmol/L)
Voice Guide
Your blood glucose is (number)
milligrams per deciliter.
Shake the control solution vial thoroughly before use. Squeeze out
the first drop and wipe it off, then squeeze out another drop and place
it on the tip of the vial cap. Hold the meter to move the absorbent hole
of the test strip to touch the drop. Once the confirmation window fills
completely, the meter will begin counting down.
3. Apply control solution
Normal
118~160