Out-of-range results
If test results fall outside the range printed on the test strip vial,
check the section of “Problem in Operation” in troubleshooting
guide and repeat the test. If you continue to get out-of-range re-
sults, it means that the system may not be working properly. Do
NOT test your blood. Please contact your local customer service
for help.
M
S
M
S
27
Doing a Control Solution Test
FIRST TAKE A TEST STRIP OUT WITH CLEAN AND DRY
HANDS.
STEP 1-
Insert test strip.
Insert a test strip with contact bars
end first and facing up into the test
slot. The monitor turns on automat-
ically and displays the following in
sequence:
“CHK” and “
”
STEP 2-
Press the M button
While the “ ” symbol appears on
the display, press the
M
button
repeatedly (three times) and then
“QC”
will appear on the display.
With the
“QC”
sign on the
display, the monitor will not store
your test result in its memory. If you
decide not to perform a control
solution test, press the
M
button
again, and the
“QC”
sign will dis-
appear.
then ambient
“ ”
and flashing
temperature
WARNING
Contact bars must be inserted all the way into the monitor or
1.
you may get an inaccurate test result.
Every time you perform a control solution test, you must en-
2.
ter into the
“QC”
test mode so that the test result will not be
stored in the monitor memory. Failure to do so will confuse the
blood glucose test result with the control solution test result in
the memory of the meter.
Summary of Contents for KIPTRACK 2 in1
Page 1: ...Blood Glucose Blood Pressure Monitoring System ACCURATE SMART PRECISE...
Page 13: ...13 Air Tube Arm Cuff Air Plug Air Jack Air Plug connect to the air jack...
Page 20: ......
Page 21: ......
Page 24: ...24 KIPTRACK Step 2 Press M button repeatedly to switch between General AC PC and QC mode...
Page 28: ......
Page 41: ......
Page 47: ...47 Error Message...
Page 52: ......
Page 57: ...57...
Page 58: ......
Page 59: ...59...
Page 60: ...60 KIPTRACK...
Page 61: ...61...
Page 62: ...62 KIPTRACK...
Page 63: ...63...