RSA-G2 Getting Started Guide
Page 47
Connecting the Test Station Power Cable
CAUTION: Before plugging the power cable into the wall outlet, make sure the instrument is com
-
patible with the line voltage. Check the label on the back of the unit to verify the voltage.
MISE EN GARDE: Avant de brancher le câble d'alimentation dans la prise murale, assurez-vous
que l'instrument est compatible avec la tension de la ligne. Consultez l'étiquette au dos de l'appareil
pour vérifier la tension.
Once all cables and lines have been connected to the modular power supplies and the RSA-G2 test station,
you can connect the power cable.
Install the power cable as follows:
1
Make sure the power switch on the lower power supply module is in the Off (O) position.
2
Plug the large, multi-pin cable from the lower power supply to the appropriate connection on the rear
of the RSA-G2 test station, as shown in
3
Connect the grounding wire to the cabinet as shown in
.
4
Plug the power cord from the lower power supply module into the wall outlet.
Figure 29
Power connection on the rear of the RSA-G2 test station.
Grounding wire
Power connection
from lower power
supply
Summary of Contents for RSA-G2
Page 1: ......
Page 9: ...Forced Convection Oven with tension clamp Technology 8...
Page 13: ...IMMERSION TESTING I CLAMPING SYSTEMS...
Page 15: ...DETA I DIELECTRIC THERMAL ANALYSIS ACCESSORY DETA...
Page 29: ...tainstruments com 2019 TA Instruments All rights reserved L50027 001...
Page 30: ...Verso Filler Page z z...
Page 31: ...n getting Started guide Revision D Issued July 2022...
Page 99: ...Issued November 2013 Supplement to Getting Started Guide TA Instruments Manual Supplement...
Page 100: ...Page 2 TA Instruments Manual Supplement This page is intentionally blank...
Page 113: ...Page 1 Page 1 RSA G2 Solids Analyzer Revision G Issued May 2022 Site Preparation Guide...