USER MANUAL T-ONE Medi/Medi-Pro MNPG06-03 Pag. 29
M04 • EMS 2 seconds slope
.
EMS 2 sec
.
slope
: to set EMS program with 2 second impulse slope.
Settable values chart
PARAMETER DESCRIPTION
MIN
MAX
Contraction frequency
20 Hertz
120 Hertz
Contraction time
4 seconds
10 seconds
Recovery frequency
1 Hertz
12 Hertz
Recovery time
2 seconds
20 seconds
Impulse width
60 microseconds 450 microseconds
Total time
1 minute
60 minutes
M05 • EMS 3 seconds slope
.
EMS 3 sec. slope
.: to set EMS program with 3 second impulse slope.
Settable values chart
PARAMETER DESCRIPTION
MIN
MAX
Contraction frequency
20 Hertz
120 Hertz
Contraction time
6 seconds
10 seconds
Recovery frequency
1 Hertz
12 Hertz
Recovery time
2 seconds
20 seconds
Impulse width
60 microseconds 450 microseconds
Total time
1 minute
60 inutes
14. SET UP
Selecting the language and contrast level on the display.
Turn T-ONE on using the button
and immediately after keeping pressed the
button
until the language list appears on the display. Release the button: now use the mouse to
move the cursors and choose the desired language or the level of contrast on the display.
Select the chosen language by using the
buttons.
Adjust the level of contrast on the display by using the buttons.
Press
to confirm your selections.