Instrument Operating Manual
CyFlow® Cube 8 Rev-27_2015-07-02
Sysmex Partec GmbH
Am Flugplatz 13, 02828 Görlitz, Germany · Phone +49 3581 8746-0 · Fax +49 3581 8746-70 · [email protected] ·
www.sysmex-partec.com
19 / 43
Schematic overview of the acquisition process for all measure modes
Green and red colored bars refer to particle count indicator of CyView.
Please note:
In the measure mode
“VC of Volume” or “Cells in Regions” there is no cleaning
cycle. Subsequently the data can be stored
. In the measure modes “VC electrodes” and “Analyze
all” the cleaning cycle will occur automatically as soon as the sample is finished.
No cleaning cycle will be
started if the measurement is terminated by the “end” button.
Beware:
for all measure modes the sample analysis automatically stops when the sample is empty
(the stop electrode is reached) even if the selected end criterion is not yet realized.
4.1.7
Measure modes
The following measure modes can be selected prior to start of an analysis:
Analyze all
Volumetric counting (VC) of volume
Volumetric counting (VC) electrodes
Cells in region