17
2. Head without rod
Apply the Stop Sleeve over the polyaxial head. Then apply the
Remobilizing Instrument as described before.
Note:
If the head has to be removed, remove nut and sleeve
using the Socket Wrench. Remove the rods. Apply the Remobi-
lizing Instrument as described above without inserting the Stop
Sleeve. This is how the locking ring will be completely removed.
Then remove the polyaxial head with the Screw Holder.
Note:
If the use of the Remobilizing instrument is hindered by
bone touching the polyaxial screw head, use the Hollow Reamer,
guided by the Screw Holder, to remove excessive bone first.
0X6.000.216_AC_0X6.000.216_AC 21.09.10 08:44 Seite 17