8
Remobilization and/or removal
Required instruments
Socket Wrench, bihexagonal,
⭋
11.0 mm, self-holding,
with straight handle 03.607.008
USS II Polyaxial Remobilizing Instrument 03.607.012
Stop Sleeve for Remobilizing without Rod 03.607.013
Hollow Reamer
⭋
12.6 mm for USS II Polyaxial 03.607.014
USS II Polyaxial Screw holder 03.607.005
Note:
Always apply the Screw Holder as a guide.
In the following situations, the USS II Polyaxial heads can be
remobilized with the Remobilizing Instrument:
1. Head with rod introduced
Loosen the nut with the Socket Wrench as far as possible.
Then slide the Remobilizing Instrument over the screw
head (make sure the red mark on the shaft with the T-Handle
is visible), and push the outer sleeve down. Turn the T-Handle
until it stops. The head is now mobile again.
16
Synthes
USS II Polyaxial Technique Guide
0X6.000.216_AC_0X6.000.216_AC 21.09.10 08:44 Seite 16