background image

SM_708378 AC 

page  3 / 3

Special instructions for mechanical cleaning

 

– In case of extreme soiling and encrusting (e.g. coagulated blood or secretion 

residues), it may be necessary to further clean the SynFrame Light-transmitting 
Rod manually.

 

Remove dirt residues on optical surfaces (see figure 1, 

 and 

) using cotton 

soaked in alcohol (70% ethanol) or a neutral cleaning agent.

 

– Check the water quality regularly to avoid formation of residues and corrosion.

 

– Do not use grease or washing agents; there may be problems with respect to 

compatibility with plastics or adhesives and compatible accessories (e.g. electri-
cal wires).

Sterilisation

Before sterilisation, make sure the SynFrame Light-transmitting Rod and the optical 

surfaces  in  particular  (see  figure  1, 

 and 

) are clean and that testing the 

 SynFrame Light-transmitting Rod does not lead to any findings that impose restric

-

tions on the use.

Further detailed instructions for reprocessing reusable devices are described in the 
Synthes brochure “Important Information”. Assembly and disassembly instructions 
of instruments “Dismantling Multipart Instruments” can be downloaded from: 
http://emea.depuysynthes.com/hcp/reprocessing-care-maintenance

Checking the fiber optics

 

Hold one side of the fiber optic (e.g. the distal end) in the direction of a bright 

ceiling lamp. For this test, do not use any cold light source. View the other side 

(light cable connection) holding it relatively close to the eye. The individual fibers 

now appear to be bright. Move the side held against the lamp. The brightness 

of the fibers now changes. If certain fibers remain dark, this is not a cause for 

concern. The intensity of illumination of the SynFrame Light-transmitting Rod 

decreases with an increasing rupture rate of the fibers.

 

– The surfaces of the light inlets and outlets must be smooth and clean. If the 

surfaces show certain deposit layers, or if rough fibers can be felt or are with

-

drawn, illumination might be inadequate. If the SynFrame Light-transmitting Rod 
is used or prepared in this condition, it is likely to be further damaged. 

Send the SynFrame Light-transmitting Rod to the manufacturer for an inspection if 

the fiber optics is damaged.

Trouble shooting

Defect

Possible cause

Possible solution

Too little 
 illumination

 

dirty fiber-optic surfaces 

 

(figure 1, 

 and 

)

 

– stubborn residue, encrusting 

on the fiber-optic surfaces

 

– wrong light cable connection

 

defective fiber optic

 

– defective light cable or light 

source

 

clean the fiber-optic surfaces 

as per instructions (manual 
cleaning)

 

– remove residues as per instruc-

tions/ check water quality

 

– check whether light cable 

 connection sits well and is 
 connected  properly

 

check fiber optic as per instruc

-

tion

 

– check light cable connection 

and light source

Yellowish 
light

 

dirty fiber optic

 

– dirty or defective light cable 

connection

 

clean fiber-optic surfaces 
 (figure 1, 

 and 

). If re-

quired, send the SynFrame 
Light-transmitting Rod in for 
servicing

 

– check light cable connection 

(e.g. by illuminating a white 
surface)

Corrosion, 
formation 
of patches, 
decoloring

 

– inadequate cleaning  

(e.g. protein residue)

 

– inadequate rinsing of the 

SynFrame Light-transmitting 
Rod between different 
preparation phases (espe-
cially before sterilisation)

 

– high chloride concentration

 

– heavy metal ions and/or sili-

cates, increased content of 
iron, copper manganese in 
water or sterilisation steam

 

– high concentration of mine-

ral substances (e.g. calcium) 
or organic substances

 

– infected or too frequently 

used disinfection or cleaning 
solutions

 

– outside rust (e.g. through 

steam or preparation along 
with damaged or rust-prone 
instruments)

 

– contact corrosion

 

– subsequent cleaning up, if 

 required by thorough rubbing

 

– ensure adequate rinsing 

 between the individual pre-
paration phases

 

– check water quality

 

– check water quality, only use 

de-ionised (distilled) water

 

– check water quality, only use 

de-ionised (distilled) water

 

– regularly replace the disinfec-

tion and cleaning solutions

 

– check maintenance systems;  

in case of preparation with 
other materials, check for 
 material  compati bility,  existing 
damages and avoid mutual 
contact

 

– avoid mutual contact with 

 other metal components