Electromagnetic Compliance
PB81499EN
84
Electromagnetic Immunity
Immunity Test
IEC 60601 Test Level
Compliance
IEC 61000-4-2: Electrostatic
Discharge (ESD)
± 6 kV contact ± 8 kV air
±2, 4, 6 kV Contact
±2, 4, 6, 8 kV Air
IEC 61000-4-4: Electrical Fast
Transient/Burst
± 2 kV for power supply lines
±2 kV
± 1 kV for input/output lines
N/A
IEC 61000-4-5: Surge
± 1 kV differential Mode
1.0 DM
± 2 kV common mode
2.0 CM
IEC 61000-4-8: Power
Frequency (50/60 Hz) Magnetic
Field
3 A/m
3 A/m
IEC 61000-4-11: Voltage Dips,
Short Interruptions and Voltage
Variations on Power Supply
Input Lines
<5 %
U
T
(>95 % dip in
U
T
) for
0.5 cycle
<5 %
U
T
(>95 % dip in
U
T
) for
0.5 cycle
40%
U
T
(60% dip in
U
T
) for
5 cycles
40%
U
T
(60% dip in
U
T
) for
5 cycles
70%
U
T
(30% dip in
U
T
) for
25 cycles
70%
U
T
(30% dip in
U
T
) for
25 cycles
5%
U
T
(95% dip in
U
T
) for
5 cycles
5%
U
T
(95% dip in
U
T
) for
5 cycles
●
Floors should be wood, concrete or ceramic tile. If floors are covered with
synthetic material, the relative humidity should be at least 30%.
●
Mains power quality should be that of a typical commercial or hospital
environment.
●
If the VelaShape III system is required to operate during power interruptions,
the system should be powered from an uninterruptible power supply.
●
Power frequency magnetic fields should be at levels characteristic of a typical
location in a typical commercial or hospital environment.
Note
U
T
is the AC mains voltage prior to application of the test level.