This chapter will assist you to operate the system for the first time, and will guide
you through the initial setup procedures needed to perform in order to start
operating the system.
Activating the System
1.
Before turning the system on, verify that the power cable is connected to the
power connection port on the system's rear panel and to the wall power outlet,
connect the umbilical cable to its connection port and ensure that the
applicator is connected to the umbilical cable.
Warning
High voltage is present inside the system. Always be aware of the
possible dangers and take proper safeguards as described in this
manual.
2.
Toggle the main power switch (on the service panel) to the
On
position; the
system initiates an automatic vacuum and RF test. When this test is
completed the system is ready for operation.
Automatic Vacuum and RF Test
The VelaShape III is equipped with an automatic vacuum and RF self-test
routine.
The test is performed after the applicator is connected.
During the test the VelaShape III system is tested for radiofrequency and vacuum
performance. During this test, the trigger or footswitch of the applicator is
disabled.
The test is comprised of three consecutive vacuum pulses, and can take up to
3 seconds.
It is important to verify that the applicator's chamber is not blocked or positioned
against a sealing surface during the test (i.e., verify that the applicator is not
applied to the patient during the test).
If a fault situation is detected, it will be displayed as an error message on the
control panel (refer to Chapter 8 –
Troubleshooting)
.