VelaShape III User Manual Operating Instructions
PB81499EN
33
Setting Treatment Parameters – Tablet Control Panel
After activation (main On/Off switch), the system starts the automatic vacuum
and RF self-test routine while displaying a startup animation. When complete, the
Main Menu
screen will be displayed (see Figure 16).
Figure 16: Main Menu Screen
The numbered tags in Figure 16 correlate to the following numbered explanations
and definitions of the screens elements.
1.
Actual Skin Temperature
– this is a reading of the patient skin's current
temperature as measured by the applicator's temperature sensor, in degrees
Celsius.
2.
Target Skin Temperature
– this is the target temperature of the skin in
degrees Celsius, set by the operator using the slide control.
3.
Slide Control
– this control sets the target temperature of the skin. Use your
finger to 'slide' the control up or down to achieve the desired target
temperature (40-45°C).
4.
RF On/Off Indicator
– RF is
On
when the ring is green and blinking; RF is
Off
when the ring is black.
5.
Session Timer
– this clock keeps track of the duration of the patient's
treatment session.
6.
Cover Type Indicator
– this tag indicates which type of cover is connected
to the applicator.
7.
Options
– use your finger to 'pull' this tab up; this will display the
Treatment
Settings
ribbon (see Figure 17).