74
TimeAnalyzer 7x00 User’s Guide
098-00208-000 Revision A – February, 2010
Chapter 3 Configuring
Recovering From Reset to TP5000 Factory Defaults
Setting Up Telnet Firewall
The firewall for Telnet must be set to “Allow” in order to initiate Telnet sessions with
the TP5000 probe module. The TimeAnalyzer 7x00 is shipped with the firewall for
Telnet for the TP5000 module set to “Allow”. The default condition for the Telnet
firewall for TP5000 is “Block”.
With CLI
Use the “show firewall” command to check if the firewall for Telnet is still set to
“Allow” for the TP5000 module. If it is set to “Block”, use the “set firewall imc telnet
allow” command to reset the firewall to “Allow”.
Example:
tp5000> set firewall imc telnet allow
With TimeMonitor PDV
Use the following procedure to set the TP5000 probe module’s Telnet firewall
to”Allow”:
1. Select the radio button for the Command mode.
2.
Click the Login button to login to the TP5000 probe module. The TP5000 probe will
prompt for Username and Password.
3. Click the Firewall Off button in order to allow for Telnet communications with the
TP5000 probe module..
Provisioning TOD Source for TP5000 Probe Module
The default TOD source for the TP5000 probe module in the TimeAnalyzer 7x00 is
System, with the Rb oscillator in free-run mode. The default TOD source for the
TP5000 module is GPS. Use the "set tod-source sys " command to set the TOD
source for the TP5000 probe module to System.
"Type "set tod-source sys "
Example:
For a duration of 500 seconds on Port 1 of IOC1: