TimeAnalyzer 7x00 User’s Guide
098-00208-000 Revision A – February, 2010
Chapter 3 Configuring
Configuring TP5000 Module Using TimeMonitor PDV
4. Enter the command
set ip imc address 192.168.106.16 netmask
192.168.106.255 gateway 255.255.255.0
and press
Enter
.
5. Enter the command
set ip imc restart
and press
Enter
.
6. Follow the steps in
, on page 56 to change the IP
address that TimeMonitor PDV uses for connecting to the TP5000 probe module.
Connecting to the TP5000 Probe MGMT Port with PDV
Use the following procedure to connect to the MGMT port (IMC card) on the TP5000
probe module using TimeMonitor PDV. The IP address for the MGMT port of the
probe module must have already been configured with the procedure in the section
in order to successfully connect to the TP5000
module with this procedure.
1. Make sure that the radio button for the
Disconnect
mode is selected.
2. Select the radio button for the desired type of connection, RS-232 or TCP/IP.
3. Select the radio button for the Command mode.
4. Click the
Login
button to login to the TP5000 probe module in the TimeAnalyzer.
5. Enter the Username and Password for the TP5000 probe module when
prompted in the main display area of the screen.
Configuring the TP5000 Probe Module for PDV Measurement
Use the following procedure to configure TimeMonitor PDV and the TP5000 probe
module for PDV measurements: