CHAPTER ONE
1-6
bc635VME/bc350VXI Time and Frequency Processor (Rev. E)
Symmetricom Inc
1.5.1.7 DAYS
When a time source signal is not present at board power up, the board will begin counting at day
000. The TFP can be operated to count days in two modes. For the default Day 000 Invalid
Mode, the TFP will not accept an input day of 000. Table 1 shows the possible combinations of
input source data and current board state on the left side, and the result of the day rollover on the
right side. Note that the table includes such combinations as where the board is set to a non-leap
year, but the source is in a leap year.
Table 1 Day 000 Invalid Mode
Combination
number
Board
year
Input mode
Source
Year
Source day
Board year
Board day
Notes
1.1.1
99
Timecode
N/A
000
99
Freerun
lost track
1.1.2
99
Timecode
99
365 – 001
99 - 00
365 – 001
1.1.3
99
Timecode
00
366 – 001
99 - 00
366 – 001
1.2.1
99
Freerun
N/A
365
99 - 00
365 – 001
1.2.2
99
Freerun
N/A
366
99 - 00
366 – 001
2.1.1
00
Timecode
00
365 – 366
00
365 – 366
2.1.2
00
Timecode
99
365 – 001
00
365 – 001
1
2.1.3
00
Timecode
00
366 – 001
00 - 01
366 – 001
2.2.1
00
Freerun
N/A
365
00
365 – 366
2.2.2
00
Freerun
N/A
366
01
366 – 001
Note 1: Day went to 366 for about one second, then went to day 001
For the optional Accept Day 000 Mode, the TFP will accept an input source with an input day of
000. Table 2 shows the possible combinations for this mode.
Table 2 Accept Day 000 Mode
Combination
number
Board
year
Input mode
Input Year
Input day
Board year
Board day
Notes
3.1.1
99
Timecode
N/A
000
99
000 – 001
3.1.2
99
Timecode
99
364 – 365
99
364 – 365
3.1.3
99
Timecode
99
365 – 001
99 – 00
365 – 001
2
3.1.4
99
Timecode
00
365 – 366
99 – 00
365 – 366
2
3.1.5
99
Timecode
00
366 – 001
99 – 00
366 – 001
2
3.2.1
99
Freerun
N/A
000
99
000 – 001
3.2.2
99
Freerun
N/A
364
99
364 – 365
3.2.3
99
Freerun
N/A
365
99 – 00
365 – 000
3.2.4
99
Freerun
N/A
366
99 – 00
366 – 000
4.1.1
00
Timecode
N/A
000 – 001
00
000 – 001
4.1.2
00
Timecode
00
365 – 366
00
365 – 366
2
4.1.3
00
Timecode
00
366 – 001
00 – 01
366 – 001
2
4.1.4
00
Timecode
99
365 – 001
00
365 – 001
2
4.2.1
00
Freerun
N/A
000
00
000 – 001
4.2.2
00
Freerun
N/A
365
00
365 – 366
4.2.3
00
Freerun
N/A
366
00 - 01
366 – 367
Note 2: Day went to 000 for about one second, then went to day 001