Shift Test
This test involves moving a weight off the center point of the top
plate to check for continued accuracy.
1. Place 5.00 kg (15.00 lb.) of weight in position 1 on the Top
Plate.
2. Move the weight to position 2 on the Top Plate. The display
should show 5.00 ± 0.005 kg (15.00 ± 0.01 lb.).
3. Repeat step 2 for positions 3, 4, and 5.
4. Move the weight to position 1 again.
5. Remove all weights. The display should read 0.000 ± 0.000kg
(0.00 ± 0.00 lb.).
6. Raise and hold up the Top Plate. Record the Cal and PAr val-
ues shown on the display.