Do NOT remove any weight from the Top Plate.
Over-Capacity Test
This test checks for the proper indication from the scale when too
much weight is placed on the top plate.
Note:
This test must immediately follow the increasing load
test; do not remove any of the weights prior to running
this test.
Place additional weight on the center of the top plate as shown in
the following chart. Use the weight that corresponds to the 7875
weight feature. The display shows a series of dashes to indicate an
over-capacity condition.
2
9.995 kg
2.5 kg
0.1 kg
2.5 ± 0.00 kg
13.995 kg
2.5 kg
0.1 kg
2.5 ± 0.00 kg
30.0 lb.
5.0 lb.
0.2 lb.
5.0 ± 0.00 lb.
3
9.995 kg
2.5 kg
5.0 ± 0.005 kg
13.995 kg
4.5 kg
7.0 ± 0.005 kg
30.0 lb.
5.0 lb.
10.0 ± 0.01 lb.
4
9.995 kg
2.5 kg
7.5 ± 0.005 kg
13.995 kg
3.0 kg
10.0 ± 0.005 kg
30.0 lb.
10.0 lb.
20.0 ± 0.01 lb.
5
9.995 kg
2.495 kg
9.995 ± 0.005 kg
13.995 kg
3.995 kg
13.995 ± 0.005 kg
30.0 lb.
10.0 lb.
30.0 ± 0.01 lb.
Step
Weight
Feature
Add
Weight
Remove
Weight
Display Result
1
9.995 kg
0.04 kg
--.---
13.995 kg
0.04 kg
--.---
30.0 lb.
0.08 lb.
--.---
Step
Weight
Feature
Add Weight
Remove
Weight