SycoTec 2003 1990 Instructions For Use Manual Download Page 3

 

Contents 

 

 

EN 

 

Contents 

1

 

User information  _____________________________________________________________________________  4

 

1.1

 

Target group of the document _______________________________________________________________  4

 

1.2

 

Applied symbols  _________________________________________________________________________  4

 

2

 

Safety  _____________________________________________________________________________________  4

 

2.1

 

Reporting obligation  ______________________________________________________________________  5

 

2.2

 

Warning signs ___________________________________________________________________________  5

 

2.3

 

Risk of infection __________________________________________________________________________  5

 

2.4

 

Magnetic fields  __________________________________________________________________________  5

 

2.5

 

Risk of burns ____________________________________________________________________________  5

 

2.6

 

Technical condition _______________________________________________________________________  6

 

3

 

Product description  ___________________________________________________________________________  6

 

3.1

 

Scope of supply __________________________________________________________________________  6

 

3.2

 

Intended purpose  ________________________________________________________________________  7

 

3.3

 

Technical specifications 

– headphones ________________________________________________________  8

 

3.4

 

Technical specifications 

– suction system ______________________________________________________  8

 

3.5

 

Operating conditions ______________________________________________________________________  8

 

3.6

 

Storage and transport conditions  ____________________________________________________________  8

 

4

 

Commissioning  ______________________________________________________________________________  9

 

5

 

Reprocessing ________________________________________________________________________________  9

 

5.1

 

Steps in the place of use __________________________________________________________________  10

 

5.2

 

Manual cleaning  ________________________________________________________________________  10

 

5.3

 

Manual disinfection ______________________________________________________________________  10

 

5.4

 

Thermal disinfection  _____________________________________________________________________  11

 

5.5

 

Sterilization in accordance with ISO 17665 ____________________________________________________  12

 

5.6

 

Storage _______________________________________________________________________________  13

 

6

 

Operation and control  ________________________________________________________________________  13

 

6.1

 

Positioning the headphones  _______________________________________________________________  13

 

6.2

 

Connecting the suction cannula to the cannula adapter __________________________________________  14

 

6.3

 

Connecting the cannula adapter to the existing suction hose of the assistant element ___________________  14

 

6.4

 

Connecting the aerosol suction to the headphones  _____________________________________________  14

 

6.5

 

Positioning the suction cannula on the patient’s mouth ___________________________________________  14

 

6.6

 

Disassembly ___________________________________________________________________________  14

 

7

 

Care, maintenance and service _________________________________________________________________  15

 

7.1

 

Troubleshooting  ________________________________________________________________________  15

 

7.2

 

Servicing ______________________________________________________________________________  15

 

8

 

Disposal ___________________________________________________________________________________  15

 

9

 

Accessories and spare parts ___________________________________________________________________  16

 

10

 

Manufacturer's warranty  ______________________________________________________________________  16

 

 

 

Summary of Contents for 2003 1990

Page 1: ...SycoTec Aerosol suction system Instructions for Use EN www sycotec eu Aerosol suction system aer x REF 2 003 1990 Copyright 2021 SycoTec All rights reserved Doc no 2 003 1812 Rev 2021 08 02...

Page 2: ...oduct may be modified or improved as a result of technical developments This does not give entitlement to have already existing products upgraded Manufacturer Distributed by SycoTec GmbH Co KG SycoTec...

Page 3: ...processing________________________________________________________________________________ 9 5 1 Steps in the place of use__________________________________________________________________ 10 5 2 Manu...

Page 4: ...date Atmospheric pressure limitation USA Professional use only Handle with care Humidity limitation Wear protective gloves Keep dry Temperature limitation Wear eye protection Stacking limitation Steam...

Page 5: ...l The suction cannulas and cannula holder must be thermally disinfected and or sterilized in autoclaves at a temperature of up to 135 C each time a patient is treated Clean and disinfect the suction h...

Page 6: ...upply Fully assembled headphones Cannula holder right Cannula holder left Suction cannulas 5 pieces Reducing adapter to 11 mm Fixing strap Disposable protective covers 100 pieces Headphone holder USB...

Page 7: ...ot be used on patients with pacemakers or implants that are sensitive to magnets due to the magnet on the headphones and adapter Patients with hearing aids Patients with skull deformities Treatments f...

Page 8: ...power suction power should be 300 l min 3 5 Operating conditions Only suitable for use in enclosed spaces Only suitable for professional use May only be used by trained staff Cannot be used as a subs...

Page 9: ...using it for the first time and reprocess it after treating each patient The aerosol suction should be reprocessed as soon as possible following treatment Also follow the processing steps in the user...

Page 10: ...of brush heads 2 Bristle material nylon Brush head length 25 and 35 mm Bristle length 5 and 10 mm Handle length min 170 mm Water of drinking water quality 30 C 5 C 86 F 41 F Take the following steps...

Page 11: ...aning and drying procedures Make sure that the medical device is dry following reprocessing In some countries national guidelines require mechanical reprocessing to be conducted using a validated wash...

Page 12: ...en operating the steam sterilizer always follow the manufacturer s instructions Observe all process parameters The suction cannulas and cannula adapters can be sterilized in a steam sterilizer B and N...

Page 13: ...torage Store the reprocessed aerosol suction in a dry dark and cool room that is free from germs and dust Please observe the sterilized product s use by date 6 Operation and control The aerosol suctio...

Page 14: ...to the cannula adapter and suction hose To do this use the magnetic fasteners on the headphones left and right side and cannula adapter 6 5 Positioning the suction cannula on the patient s mouth 1 Po...

Page 15: ...ervice partners Unauthorized repairs or modifications to the product invalidate the product s license and warranty Please contact your dealer for information about authorized service partners 8 Dispos...

Page 16: ...oloration e g of plastic parts Warranty claims are only valid if all the instructions in the product s user manual are followed SycoTec does not assume any liability in the event of gross negligence o...

Reviews: