SycoTec 2003 1990 Instructions For Use Manual Download Page 10

 

5 Reprocessing 

 

 

10 

EN 

 

5.1 

Steps in the place of use 

Preparation at the place of use: 

 

 

Wear protective gloves 

 

Wear a mask 

 

Wear eye protection 

 

Wear protective clothing 

 

Take the following steps immediately after treating each patient: 

 

Remove the protective covers from the headphones. 

 

Remove any heavy soiling such as residual cement, composite or blood from the components of the aerosol suction 
immediately in the place of use. 

 

You can use the accessories described in chapters 5.2 and 5.3 for this purpose. 

 

Only the aerosol suction needs to be reprocessed. 

 

The headphones only need to be cleaned using wipe-down disinfection. 

 

These steps must be taken before the aerosol suction is reprocessed. 
These steps must be performed in addition to normal practice hygiene. 
 

5.2 

Manual cleaning 

Clean the surfaces of the suction cannulas and cannula adapters. Use the following materials: 

 

  Brush, e.g. cleaning brush with nylon bristles (double-sided) 

 

– Number of brush heads: 2 

 

– Bristle material: nylon 

 

– Brush head length: 25 and 35 mm 

 

– Bristle length: 5 and 10 mm 

 

– Handle length: min. 170 mm 

  Water of drinking-water quality 30°C ± 5°C (86°F ± 41°F) 

 

Take the following steps: 

  1.  Clean the suction cannula and cannula adapter under running 

water with continuous and oscillating brushing motions. 

  2.  Perform a visual inspection of the surfaces. Repeat step 1 if any 

dirt remains on the suction cannula. 

  3.  Then perform manual disinfection. 

 

 

5.3 

Manual disinfection 

Disinfect the surfaces of the suction cannulas and cannula adapters. 
Use the following materials: 

 

Disinfectant solution and cotton cloth (e.g. SciCan OPTIM Blue / Metrex CaviCide / DÜRR DENTAL FD 322) 

 

or 

 

Disinfectant wipes (e.g. SciCan OPTIM Blue Wipes / Metrex CaviWipes / Dr. Schumacher CLEANISEPT WIPES 
FORTE) 

 

or 

 

Place the suction cannulas and cannula adapters in a disinfection bath (non-fixing and aldehyde free, all parts must 
be covered). 

 

 

Observe the exposure times of the cleaning agents and disinfectants 

 

 

Do 

not

 put the 

headphones

 into a disinfection bath. 

 
The manufacturer has tested the disinfectants to ensure that they are compatible with the aerosol suction. 
 

 

Summary of Contents for 2003 1990

Page 1: ...SycoTec Aerosol suction system Instructions for Use EN www sycotec eu Aerosol suction system aer x REF 2 003 1990 Copyright 2021 SycoTec All rights reserved Doc no 2 003 1812 Rev 2021 08 02...

Page 2: ...oduct may be modified or improved as a result of technical developments This does not give entitlement to have already existing products upgraded Manufacturer Distributed by SycoTec GmbH Co KG SycoTec...

Page 3: ...processing________________________________________________________________________________ 9 5 1 Steps in the place of use__________________________________________________________________ 10 5 2 Manu...

Page 4: ...date Atmospheric pressure limitation USA Professional use only Handle with care Humidity limitation Wear protective gloves Keep dry Temperature limitation Wear eye protection Stacking limitation Steam...

Page 5: ...l The suction cannulas and cannula holder must be thermally disinfected and or sterilized in autoclaves at a temperature of up to 135 C each time a patient is treated Clean and disinfect the suction h...

Page 6: ...upply Fully assembled headphones Cannula holder right Cannula holder left Suction cannulas 5 pieces Reducing adapter to 11 mm Fixing strap Disposable protective covers 100 pieces Headphone holder USB...

Page 7: ...ot be used on patients with pacemakers or implants that are sensitive to magnets due to the magnet on the headphones and adapter Patients with hearing aids Patients with skull deformities Treatments f...

Page 8: ...power suction power should be 300 l min 3 5 Operating conditions Only suitable for use in enclosed spaces Only suitable for professional use May only be used by trained staff Cannot be used as a subs...

Page 9: ...using it for the first time and reprocess it after treating each patient The aerosol suction should be reprocessed as soon as possible following treatment Also follow the processing steps in the user...

Page 10: ...of brush heads 2 Bristle material nylon Brush head length 25 and 35 mm Bristle length 5 and 10 mm Handle length min 170 mm Water of drinking water quality 30 C 5 C 86 F 41 F Take the following steps...

Page 11: ...aning and drying procedures Make sure that the medical device is dry following reprocessing In some countries national guidelines require mechanical reprocessing to be conducted using a validated wash...

Page 12: ...en operating the steam sterilizer always follow the manufacturer s instructions Observe all process parameters The suction cannulas and cannula adapters can be sterilized in a steam sterilizer B and N...

Page 13: ...torage Store the reprocessed aerosol suction in a dry dark and cool room that is free from germs and dust Please observe the sterilized product s use by date 6 Operation and control The aerosol suctio...

Page 14: ...to the cannula adapter and suction hose To do this use the magnetic fasteners on the headphones left and right side and cannula adapter 6 5 Positioning the suction cannula on the patient s mouth 1 Po...

Page 15: ...ervice partners Unauthorized repairs or modifications to the product invalidate the product s license and warranty Please contact your dealer for information about authorized service partners 8 Dispos...

Page 16: ...oloration e g of plastic parts Warranty claims are only valid if all the instructions in the product s user manual are followed SycoTec does not assume any liability in the event of gross negligence o...

Reviews: