42
A-96.250.501 / 280915
AMI Hydrazine
Maintenance
Disassemble
the flow cell
1
Switch off the instrument.
2
Stop the sample flow at the main tap before the sample inlet.
3
Open the grab sample tap [J] to empty the flow cell.
4
Remove all sensors, see
Cleaning of Hydrazine sensor, p. 38
Cleaning of Reference Electrode, p. 40
and temperature sen-
sor.
5
Put the rubber cap on the tip of the reference electrode and
plug cap on sensor plug.
6
Remove the following parts from the flow cell block [I], but with-
out however removing the flow cell block from the panel:
–
Tube from sample inlet to DIPA bottle [B]
–
Tube from DIPA bottle to hydrazine sensor [C]
–
Constant head cover [D]
–
Constant head tube [E]
–
Overflow tube to waste (long) [G]
–
Overflow tube to sensor (short) [H]
–
Grab sample tap [J]
–
Sample inlet [M] and grab sample outlet (not visible)
–
Filter vessel [P]
–
Filter [O]
–
Sample outlet [R]
–
Hall-Effect sensor [N]
–
Flow regulating valve [L]
7
Clean all acrylic parts with a soft brush (e.g. a bottle cleaner)
and soapy water. Remove calcareous deposits with a common
household deliming agent with standard concentrations.
8
Clean the bores of the flow cell block with pipe cleaners.