AMI Hydrazine
Maintenance
6.4.
Cleaning of Hydrazine sensor
1
Close the flow regulating vale [A].
2
Wait until rotor [C] stops and hydrazine reading is 0 ppm.
3
Shut off power of the instrument.
4
Disconnect the BNC connector [G] from the hydrazine sensor
[F].
Prevent it from getting wet.
5
Unscrew and remove one of the two knurled nuts [H].