AMI Hydrazine
Program List and Explanations
A-96.250.501 / 280915
63
3 Maintenance
3.1 Calibration
3.1.1
Zero Hydrazine
Zero calibration (0 ppb). See
for more de-
tails.
3.1.2
Process Hydrazine
For a process calibration of the hydrazine sensor follow the instruc-
tions on the screen. Save the value with the [Enter] key. See
3.1 Simulation
To simulate a value or a relay state, select the
alarm relay,
relay 1or 2
signal output 1or 2
with the [
] or [
] key.
Press the [Enter] key.
Change the value or state of the selected item with the [
] or
[
] key.
Press the [Enter] key.
The value is simulated by the relay/signal output.
At the absence of any key activities, the instrument will switch back
to normal mode after 20 min. If you quit the menu, all simulated val-
ues will be reset and the transmitter reboots.
3.2 Set Time
Adjust date and time.
3.4.1
3.4.2
3.4.3
3.4.4
3.4.5
Alarm Relay:
Relay 1:
Relay 2
Signal Output 1:
Signal Output 2
Open or closed.
Open or closed.
Open or closed.
Actual current in mA
Actual current in mA