13.2.1
Sensor/ADC settings for tap balancing
A dual-tap camera will usually have the “Sensor/ADC
Adjustment” feature enabled which results in a
“Luminance” panel with additional fields and controls.
Current gain balance between a left and a right sensor
tap is displayed in the right bottom field in dB where a
higher gain for the right tap results in a positive balance
value. The current (volatile) balance can be adjusted by
a “Gain balance” slider on the bottom.
A single set of gain/offset factory settings are stored in
the camera. Those settings determine the zero point for the user accessible gain and
offset values, usually called calibration values. The factory settings for those values
become available for change when clicking on the “Calibrate” button. A warning will be
issued as follows:
After confirming the warning dialog with OK a “Sensor/ADC” panel becomes available
and replaces the “Luminance” panel.
Clicking on Left and Right radio buttons or directly on
the gain/offset values makes those values accessible
for changes. They can be changed by slider or by
editing them in the edit field. Once having finished with
adjusting all values the OK button can be clicked for
making the new settings persistent in the camera.
NOTE: Factory settings for gain/offset will be irrevocably overwritten when
onfirming the final message box with OK: