![Svantek SVAN 971 User Manual Download Page 27](http://html1.mh-extra.com/html/svantek/svan-971/svan-971_user-manual_1409279027.webp)
SVAN 971 User Manual
27
After calibration measurement stop, the
New Factor
(difference between
Calibration Level and Calibration Measurement, calculated in dB) is displayed and
it will be proposed to save the new calibration factor by pressing
<Enter>
(
Enter=Conf.
), or reject it by pressing
<Esc>
. In both cases the instrument exits
the
Calibration
screen.
It is recommended to repeat calibration measurements few times. Obtained results should be almost the
same (with
0.1 dB difference). Reasons for unstable results are as follows:
•
calibrator is not properly attached to the instrument,
•
there are external acoustic disturbances such as high noise levels nearby,
•
calibrator or measurement channel (the microphone, the preamplifier or the instrument itself)
are damaged.
Note: During the calibration measurement, external disturbances (acoustic noise or
vibrations) should not exceed a value of 100 dB (when using a calibrator that generates
114 dB).
5. Press
<Enter>
to accept and save the new calibration
factor.
If calculated calibration factor is out of the ±3 dB range
the special warning appears on the screen. If the
calibration factor is out of the ±20 dB range the calibration
factor is not accepted.
<ENT>
6. Detach the calibrator from the microphone.
Note: To quit the calibration procedure without saving the calibration factor, press <ESC>.
4.2.2 History of calibrations - Last Calibration
The
Last Cal.
window displays up to ten last calibration
records and one position, that erases all calibration
records (
Clear Hist.
).
<ENT>
To erase all calibrations records in the history, choose the position
Clear Hist.
and press
<Enter>
.
The instrument requests confirmation of the selected operation.
4.2.3 Post measurement calibration
– Post Calibration
Sometimes it is required to perform so called post-
calibration of the instrument. The
Post Cal.
position
allows you to perform additional calibration after
measurement session and add its results to the data file.
There are three options for saving calibration results: not
to save (
Off
), save in the last file (
Last File
) or save in the
files which will be created after the last calibration
(
After Cal.
).
<ENT>
Summary of Contents for SVAN 971
Page 98: ...SVAN 971 User Manual 98 ...
Page 154: ...SVAN 971 User Manual Appendixes 154 1000 Hz 1250 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Page 155: ...SVAN 971 User Manual Appendixes 155 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Page 156: ...SVAN 971 User Manual Appendixes 156 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Page 157: ...SVAN 971 User Manual Appendixes 157 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Page 206: ...SVAN 971 User Manual Appendixes 206 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Page 207: ...SVAN 971 User Manual Appendixes 207 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Page 208: ...SVAN 971 User Manual Appendixes 208 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Page 209: ...SVAN 971 User Manual Appendixes 209 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...