SVAN 971 User Manual
– Appendixes
144
Periodical test upper frequency
8 kHz
Linear Operating Ranges
Two measuring ranges
are available: “LOW” and "HIGH".
The starting point at which tests of level linearity shall begin is 94.0 dB for the frequencies specifies below.
Table C.1.1.
Linear operating range:
“LOW” for the sinusoidal signal and microphone sensitivity 35 mV/Pa
[dB]
L
AS/F
L
BS/F
L
CS/F
L
ZS/F
L
AeqT
L
BeqT
L
CeqT
L
AE
L
Cpeak
(t
int
= 2 s)
from
to
from
to
from
to
from
to
from
to
from
to
from
to
from
to
from
to
31,5 Hz
25
83
25
106
25
120
30
123
25
83
25
106
25
120
28
86
50
123
500
Hz
25
119
25
122
25
123
30
123
25
119
25
122
25
123
28
122
50
126
1 kHz
25
123
25
123
25
123
30
123
25
123
25
123
25
123
28
126
50
126
4 kHz
25
124
25
122
25
122
30
123
25
124
25
122
25
122
28
127
50
126
8 kHz
25
122
25
120
25
120
30
123
25
122
25
120
25
120
28
125
50
123
12.5
kHz
25
118
25
117
25
117
30
123
25
118
25
117
25
117
28
121
50
120
Table C.1.2.
Linear operating range: „HIGH“ (primary level range) for the sinusoidal signal and
microphone sensitivity 35 mV/Pa
[dB]
L
AS/F
L
BS/F
L
CS/F
L
ZS/F
L
AeqT
L
BeqT
L
CeqT
L
AE
L
Cpeak
(t
int
= 2 s)
from
to
from
to
from
to
from
to
from
to
from
to
from
to
from
to
from
to
31,5 Hz
35
97
35
120
35
134
40
137
35
97
35
120
35
134
38
100
60
137
500
Hz
35
133
35
136
35
137
40
137
35
133
35
136
35
137
38
136
60
140
1 kHz
35
137
35
137
35
137
40
137
35
137
35
137
35
137
38
140
60
140
4 kHz
35
138
35
136
35
136
40
137
35
138
35
136
35
136
38
141
60
139
8 kHz
35
136
35
134
35
134
40
137
35
136
35
134
35
134
38
139
60
137
12.5
kHz
35
132
35
131
35
131
40
137
35
132
35
131
35
131
38
135
60
134
Note: For the signals with the crest factor n > 1.41 upper measuring range of the RMS (LEQ
and SPL) is reduced. The valid upper limit can be calculated according to the below given
formula:
, where A is the upper limit for the sinusoidal signal
Example: For the crest factor n = 10 the upper limit is A10 = 120 dB
Measuring ranges
Reference measuring range
of the acoustic pressure
Low
Measuring frequency range
of the acoustic pressure
(-3 dB) 10
Hz ÷ 20 000 Hz.
Basic measurement error
of the acoustic pressure
< 0.7 dB
(measured for the reference
conditions, see below).
2
log
20
137
n
A
n
Summary of Contents for SVAN 971
Page 98: ...SVAN 971 User Manual 98 ...
Page 154: ...SVAN 971 User Manual Appendixes 154 1000 Hz 1250 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Page 155: ...SVAN 971 User Manual Appendixes 155 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Page 156: ...SVAN 971 User Manual Appendixes 156 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Page 157: ...SVAN 971 User Manual Appendixes 157 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Page 206: ...SVAN 971 User Manual Appendixes 206 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Page 207: ...SVAN 971 User Manual Appendixes 207 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Page 208: ...SVAN 971 User Manual Appendixes 208 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Page 209: ...SVAN 971 User Manual Appendixes 209 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...