![Svantek SVAN 971 User Manual Download Page 100](http://html1.mh-extra.com/html/svantek/svan-971/svan-971_user-manual_1409279100.webp)
SVAN 971 User Manual
– Appendixes
100
#1,U971,N1234,W1.10.2,Q0.01,M1,R1,F2:1,F3:2,F1:3,F2:4,F3:5,F1:6,J2:1,J3:2,J1:3,J3:4,J3:5,J1:6,f1,
C1:1,C0:2,C2:3,C1:4,C0:5,C2:6,B0:1,B3:2,B15:3,b0,d1s,D10s,K5,L0,Y3,y0,XT0,XL100,XQ0,Xq0,XC1
15:1,XC115:2,XC115:3,Xl115:1,Xl115:2,Xl115:3,XA0,XD-1:1,XD-1:2,XD-1:3,XD-1:4,XD-1:5,XD-
1:6,S0,T1,e480,c1:1,c1:2,c1:3,h0:1,h0:2,h0:3,x3:1,x3:2,x5:3,m0,s0,l100,O10,o0,t0;
means that:
•
SVAN 971 is investigated (U971);
•
its number is 1234 (N1234);
•
software version number is 1.07.1 (W1.10.2);
•
calibration factor is equal to 0.01 dB (Q0.01);
•
LEVEL METER
is selected as the measurement function (M1);
•
range is
LOW
(R1);
•
A
filter is selected in profile 1, SLM function (F2:1);
•
C
filter is selected in profile 2, SLM function (F3:2);
•
Z
filter is chosen in profile 3, SLM function (F1:3);
•
A
filter is selected in profile 1, DOSE function (F2:4);
•
C
filter is selected in profile 2, DOSE function (F3:5);
•
Z
filter is chosen in profile 3, DOSE function (F1:6);
•
A
Peak filter is selected in profile 1, left channel, SLM function (J2:1);
•
C
Peak filter is selected in profile 2, left channel, SLM function (J3:2);
•
Z
Peak filter is selected in profile 3, left channel, SLM function (J1:3);
•
C
Peak filter is selected in profile 1, both channels, DOSE function (J3:4);
•
C
Peak filter is selected in profile 2, both channels, DOSE function (J3:5);
•
Z
Peak filter is selected in profile 3, both channels, DOSE function (J1:6);
•
Z
filter is selected for
1/1 OCTAVE
or
1/3 OCTAVE
analysis (f1)
•
FAST
detector is selected in profile 1, SLM function (C1:1);
•
IMPULSE
detector is chosen in profile 2, SLM function (C0:2);
•
SLOW
detector is selected in profile 3, SLM function (C2:3);
•
FAST
detector is selected in profile 1, DOSE function (C1:4);
•
IMPULSE
detector is chosen in profile 2, DOSE function (C0:5);
•
SLOW
detector is selected in profile 3, DOSE function (C2:6);
•
logger’s buffer is not filled by the results from profile 1 (B0:1);
•
Lpeak
and
Lmax
values are stored in the files of the logger from profile 2 (B3:2);
•
Lpeak
,
Lmax
,
Lmin
and
Leq
values are stored in the files of the logger from profile 3 (B15:3);
•
results of
1/1 OCTAVE
or
1/3 OCTAVE
analysis are not stored in the files of the logger (b0);
•
results are stored in a logger’s file every 1 second (d1s);
•
integration period is equal to 10 seconds (D10s);
•
the measurement has to be repeated 5 times (K5);
•
linear detector is selected to the
Leq
calculations (L0);
•
delay of the start of the measurements is equal to 3 seconds (Y3);
•
synchronization the start of measurement with RTC is switched off (y0);
•
logger triggering mode is switched off (XT0);
•
logger triggering level is set to 100 dB (XL100);
•
number of the records before the triggering saved in a file of the logger is equal to 0 (XQ0);
•
number of records registered, after the moment in which the measured signal does not fulfil
any longer the condition of the triggering, is equal to 0 (Xq0);
•
threshold level for PTC calculation in profile 1, is set to 115 dB (XC115:1);
•
threshold level for PTC calculation in profile 2, is set to 115 dB (XC115:2);
Summary of Contents for SVAN 971
Page 98: ...SVAN 971 User Manual 98 ...
Page 154: ...SVAN 971 User Manual Appendixes 154 1000 Hz 1250 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Page 155: ...SVAN 971 User Manual Appendixes 155 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Page 156: ...SVAN 971 User Manual Appendixes 156 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Page 157: ...SVAN 971 User Manual Appendixes 157 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Page 206: ...SVAN 971 User Manual Appendixes 206 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Page 207: ...SVAN 971 User Manual Appendixes 207 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Page 208: ...SVAN 971 User Manual Appendixes 208 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Page 209: ...SVAN 971 User Manual Appendixes 209 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...