background image

 SV 102 USER GUIDE 

© SVANTEK 
 

 
 

TABLE OF CONTENTS

TABLE OF CONTENTS

TABLE OF CONTENTS

TABLE OF CONTENTS    

 
 

INTRODUCTION..................................................... 2

 

CONTROL PUSH-BUTTONS ...................................... 3

 

BEFORE YOU TURN INSTRUMENT ON........................ 3

 

TURNING INSTRUMENT ON AND OFF ........................ 3

 

WARM UP TIME ..................................................... 4

 

BATTERY CHECK.................................................... 5

 

EXCHANGING BATTERIES ....................................... 5

 

SETTING DATE AND TIME ....................................... 5

 

MEASUREMENT SETUP ........................................... 6

 

CALIBRATION ....................................................... 8

 

MAKING A SIMPLE MEASUREMENT & USING 
DISPLAY OPTIONS ............................................... 10

 

SAVING MEASUREMENT RESULTS .......................... 12

 

LOADING MEASUREMENT RESULTS ........................ 13

 

DELETING MEASUREMENT RESULTS ....................... 13

 

 

 

 

Summary of Contents for SV 102

Page 1: ...SV 102 Dual Channel Acoustic Dosimeter QUICK START GUIDE draft www svantek com...

Page 2: ...TTONS 3 BEFORE YOU TURN INSTRUMENT ON 3 TURNING INSTRUMENT ON AND OFF 3 WARM UP TIME 4 BATTERY CHECK 5 EXCHANGING BATTERIES 5 SETTING DATE AND TIME 5 MEASUREMENT SETUP 6 CALIBRATION 8 MAKING A SIMPLE...

Page 3: ...performing binaural dose measurements and the 1 1 octave analysis simultaneously SV102 provides Type 2 accuracy The SV 102 dosimeter is offered with the dedicated SV 25D ceramic microphones with hous...

Page 4: ...E YOU YOU YOU YOU TUR TUR TUR TURN N N N INSTRUMENT INSTRUMENT INSTRUMENT INSTRUMENT ON ON ON ON Plug in the microphone s before turning on the instrument In the case of one channel operation please u...

Page 5: ...n when it happens is the exchanging of the connections of the left and right microphone If the TEDS or MEMORY factor is correct select it and press If none of the value is correct select any of them a...

Page 6: ...S EXCHANGING BATTERIES EXCHANGING BATTERIES Switch the instrument off before changing batteries Unscrew black bottom cover of the instrument e g using a coin Change the batteries follow polarity as il...

Page 7: ...SUREMENT SETUP and configure using and push buttons following parameters START DELAY period between pressing START and the beginning of measurement Press down to go to next position INTEGR PERIOD inte...

Page 8: ...time constant FAST SLOW IMPULSE Availability of filter selection A C Z depends on chosen MEASUREMENT FUNCTION for dosimeter modes filter settings in profiles are fixed for sound modes the setting of...

Page 9: ...STANDARD SETUPS OSHA MSHA DOD ACGIH ISO 85 ISO 95 by selecting proper option in FILE LOAD SETUP STANDARD SETUP window and pressing CALIBRATION CALIBRATION CALIBRATION CALIBRATION Go to MENU CALIBRATIO...

Page 10: ...calibration factor and to store it in TEDS memory of the microphone or if you don t want to accept the calibration factor press and push buttons to select NO repeat the calibration Instrument returns...

Page 11: ...B E_8h Pa2 h SEL8 dB PSEL dB PCTC PCTP RMS dB for SLM the following results are available SPL LEQ SEL Lden LEPd Ltm3 Ltm5 Lxx OVL PEAK MAX MIN You can see main measurement results by pressing right an...

Page 12: ...SV 102 USER GUIDE SVANTEK 11...

Page 13: ...h buttons or go to MENU FILE SAVE You can edit the name of the file press to edit name and press You can use SAVE NEXT option the last digit of the file s name will be automatically incremented every...

Page 14: ...to load and press DELETING MEASUREMENT RESULTS DELETING MEASUREMENT RESULTS DELETING MEASUREMENT RESULTS DELETING MEASUREMENT RESULTS To delete result or setup file one by one go to MENU FILE DELETE s...

Reviews: