![Sutter Instrument SmartShutter Lambda SC Operation Manual Download Page 41](http://html1.mh-extra.com/html/sutter-instrument/smartshutter-lambda-sc/smartshutter-lambda-sc_operation-manual_1407918041.webp)
LAMBDA SC OPERATION MANUAL – REV. 1.14C (20190606)
33
The following paragraphs describe the special commands in more detail.
10.4.1
Status
The Status command is used to return information about the state and mode of the attached
SmartShutter.
Table 10-5. Status command return codes and data
Order
Category
Byte
Num.
Value
(Decimal,
hexadecimal, &
binary)
Description
1 Command echo
(1 byte)
1
204
CC
11001100
The Status command byte code echoed back.
2 Shutter
Open/Closed State
(1 byte)
2
170
AA
10101010
Open: Shutter is in the opened state.
172
AC
10101100
Closed: Shutter is in the closed state.
3 Shutter Mode
(1 byte)
3
219
DB
11011011
Not Connected: Indicates that no
SmartShutter is connected.
220
DC
11011100
Fast: Indicates that the SmartShutter is in
fast mode.
221
DD
10111011
Soft: Indicates that the SmartShutter is in
soft mode.
222
DE
10111100
Neutral Density: Indicates that the
SmartShutter is in neutral-density mode. A
second byte follows with number of
microsteps (1 – 144).
NOTE: From this point onwards in this table, byte numbers shown in
parenthesis are true only if Shutter Mode (Byte 3) = Neutral Density (222
decimal (DE hexadecimal)).
4 Neutral Density
Microsteps
(1 byte)
(4)
1 – 144
01 – 90
00000001 - 10010000
Neutral Density Microsteps: Contains the
number of microsteps (1 through 144) for
the SmartShutter’s neutral density mode.
Note that this byte is present (and at this
position) in Status structure only if the
Shutter Mode (Byte 3) contains the value for
Neutral Density (222 decimal (DE
hexadecimal)).
Summary of Contents for SmartShutter Lambda SC
Page 3: ......
Page 4: ......
Page 10: ...2 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...
Page 12: ...4 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...
Page 14: ...6 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...
Page 16: ...8 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...
Page 18: ...10 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...
Page 22: ...14 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...
Page 30: ...22 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...
Page 32: ...24 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...
Page 66: ...58 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...
Page 70: ...62 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...
Page 72: ...64 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...
Page 74: ...66 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...
Page 76: ...68 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...
Page 88: ...80 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...