32
LAMBDA SC OPERATION MANUAL – REV. 1.14C (20190606)
Command
Value
(Decimal,
hexadecimal, &
binary)
Description
Low
FA + B2
11
10110010
Restore Controller
to the Factory-
Default
Configuration
250 + 192
FA + C0
11
11000000
Changes the controller’s configuration to that of the factory
default.
Save the Current
Configuration to
the Controller
250 + 193
FA + C1
11
11000001
Saves the current configuration to the controller. This saved
configuration will be used next time the controller is
powered up or is reset.
Set Number of
Repeat Cycles for
Free Run
250 + 240 + 2
bytes
FA + F0 + 2
bytes
11
11
2 bytes
Sets the number of repeat cycles for the Free Run feature.
3
rd
and 4
th
bytes contain the number of repeat cycles as a 16-
bit integer ranging from 0 to 65,000. Any number above
65,000 (65,001 to 65,535) sets up the Free Run for
Continuous operation (i.e., infinite repeat cycles).
Start Free Run on
Power Up
250 + 241
FA + F1
11
11110001
Starts the Free Run immediately after the controller powers
up. The number of repeat cycles does not need to be set
beforehand.
Start Free Run on
Trigger Pulse
from TTL IN
250 + 242
FA + F2
11
11110010
Starts the Free Run immediately after trigger pulse is
received on TTL IN. The number of repeat cycles must be
set beforehand.
Start Free Run
immediately
250 + 243
FA + F3
11
11110011
Starts the Free Run immediately on receiving this command.
The number of repeat cycles must be set beforehand.
Reset Controller
to the Last Saved
Configuration
251
FB
11111011
Resets the controller and sets the configuration to the one
that was last saved.
Get Controller
Type and
Configuration
253
FD
11111011
Queries the controller as to its type and configuration.
NOTE: The “TTL IN Falling Edge Triggers SmartShutter to Toggle Open/Close” command
was introduced in versions of the Lambda SC programmed with Version 1.08 or later of the
firmware. Before attempting to use this command, it is strongly advised that the version of
the firmware with which the connected Lambda SC is programmed is checked by first issuing
the “Get Controller Type and Configuration” command.
Summary of Contents for SmartShutter Lambda SC
Page 3: ......
Page 4: ......
Page 10: ...2 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...
Page 12: ...4 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...
Page 14: ...6 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...
Page 16: ...8 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...
Page 18: ...10 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...
Page 22: ...14 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...
Page 30: ...22 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...
Page 32: ...24 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...
Page 66: ...58 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...
Page 70: ...62 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...
Page 72: ...64 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...
Page 74: ...66 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...
Page 76: ...68 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...
Page 88: ...80 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...