6.3.2 Calibration Parameters (CALP)
No.
Sign
Range
Default
Description
Set
120
ZP
0~1
0
Zero Auto-tracking Permission
0: oFF
1: on
[Only after weight being stable and the
zero variation in ‘Zero Auto-tracking Time’ is
within the range of ‘Zero Auto-tracking Limit’, the
result of Zero Auto-tracking will be valid]
121
Zt
0.1~9.9
1.0
Zero Auto-tracking Time
[s]
122
Zr
0.1~50.0
1.0
Zero Auto-tracking Limit
[Division]
123
nZ
0~50000
50
Manual/Auto Zero Limit
Only when the weight variation caused by Zero
Value changing is within this range, Manual Zero
and Auto Zero will be valid.
Suggestion: Set value
≤
(Scale Capacity × 4%).
124
dZ
0~99
5
Zero Display Limit
When the absolute value of weight is within this
range, the weight display value will return to zero.
125
tL
1~99999
12000
[*]
Total Capacity of Loadcells
Total Capacity of Loadcells = Loadcell Capacity ×
Loadcell Number. Only for query.
126
SE
0.500~5.000
2.000
[*]
Output Sensitivity of Loadcell
[mV/V]
Only for query.
127
Un
0
0
Unused
128
In
0~1
0
Zero Upon Power Up
0: oFF
1: on
[without Power-down Protection]
129
It
0~1800
10
Power Up Zero Time
[s]
130
Ir
0~50000
50
Power Up Zero Limit
Suggestion: Set value
≤
(Scale Capacity × 20%).
[*]: ‘RAM Reset’ operation has no effect on the parameter.