Chapter 10 Setting
www.supmea.com
13
3.
Set the offset current of temperature, the range is±1.00mA.
4.
The filter time range is 0-120s, the low pass filter of software will active
when the current from one point to another point if user sets the filter time.
5.
Set the current 2 output mode(fixed / last) when instrument enter into HOLD
Type.
10.4 Relay 1 settings
1.
Press UP/DOWN key to ON/OFF (enable/ disable) relay1.
2.
Close set point: active point for EC/ Resistivity/Salinity/TDS.
3.
Open set point: inactive point for EC/ Resistivity/Salinity/TDS.
4.
Delay time: the range is 0-120s. Relay needs to delay first then active if the
measuring data is reach to close set point.
5.
For example: If user wants turn on the pump at EC 1000uS and turn off it at
EC 400uS, then the close S.P. needs to set to 1000uS, Open S.P. sets to
400uS.
10.5 Relay 2 settings
1.
Press UP/DOWN key to ON/OFF (enable/ disable) relay2.