10.2 Current 1 settings
1.
Set the corresponding 4.00mA to EC/ Resistivity/Salinity/TDS.
2.
Set the corresponding 20.00mA to EC/ Resistivity/Salinity/TDS, the
difference between 4.00mA and 20.00 mA at least for EC is
1.00/10.0/100(uS/mS), for Resistivity is 1.00MΩ,for Salinity is 1.00g/kg,for
TDS is 100 ppm.
3.
Set the offset current ,the range is± 1.00mA.
4.
The filter time range is 0-120 seconds, the low pass filter of software will
active when the current from one point to another point if user sets the filter
time.
5.
Set the current 1 output mode(fixed / last) when instrument enter into HOLD
Type.
10.3 Current 2 settings
1.
Set the corresponding 4.00mA to temperature.
2.
Set the corresponding 20.00mA to temperature, the difference at least
between 4.00mA and 20.00mA is 10.0
℃
.