Chapter 4 Menu window details
www.supmea.com
29
M32
Window for selecting the totaliziers' working unit
M33
Select totalizer multiplier
The multiplier ranges from 0.001 to 10000
M34
Turn on or turn off the NET totalizer
M35
Turn on or turn off the P0s totalizer
M36
Turn on or turn off the NEG totalizer
M37
(1) Totalizer reset
(2) Restore the instrument to the default parameters as the
manufacturer did
By pressing the dot key followed by the backspace key. Take care or
make note on the parameters before doing the restoration
M38
Press a key to run or to stop totalizer for easier calibration
M39
Operational interface language selection in Chinese and English. This
selection makes it possible that more than 2 billions of people on the
world can read the menu.
M40
Flow rate damper for a stable value. The input range is 0 to 999
seconds.
0 means there is no damping. Default value is 10 seconds
M41
Lower flow rate cut.offto avoid invalid accumulation.
M42
Zero point setup under the condition when there is no liquid running
inside the pipe.
M43
Clear the zero point set by the user, and restore the zero point set by
the manufacturer
M44
Set up a manual flow bias. Generally this value should be 0.
M45
Scale factor for the instrument. The default value is‘1'.
Keep this value as‘1', when no user calibration has been made.
M46
Network environment Identification Number. Any integer can be
entered except 13(0DH, carriage return), 10 (0AH, 1ine feeding), 42
(2AH), 38, 65535.
Every set of the instrument in a network environment should have a
unique IDN. Please refer to the chapter for communication.