For accurate blood pressure measurements, ensure that the circumference of
the arm fits within the range markings on the cuff.
The
SunTech 247
is not intended for continuous monitoring. Although the
blood pressure cuff and cable are defibrillator proof, the temperature probe
and SpO
2
sensor are not. Do not leave the device unattended while taking
measurements on a patient.
Only use such accessories as are recommended for use with this device. A list
of recommended accessories is on page 61.
Do not operate the
SunTech 247
near flammable anesthetics or volatile
vapors. An explosion may result.
Compressing the pneumatic tubing may cause system errors.
Do not use the device if it has failed its diagnostic self test or if it displays a
greater than zero pressure with no cuff attached or a value of functional
oxygen saturation or temperature with no sensor attached.
Prevent water or other fluids from entering any connectors or vents on the
device. Should this happen, all connectors should be dried with warm air.
Then check the calibration of the device and operating functions before
reusing.