41
Problem: Air leak, Loose cuff, or Blocked or pinched hose
Solutions:
1.
Check that the hose has no sharp bends or is pinched.
2.
Check that the patient is not lying on the cuff.
3.
Check that the cuff is in the correct position.
4.
Check that the hose is connected to the system and the cuff.
5.
Check that the cuff is properly tightened.
6.
Check that the correct size cuff is being applied.
7.
Check that the cuff is not leaking air.
8.
Check that the hose connections are not damaged or loose.
9.
Take another BP reading.
Status Codes: 800, 900, 910, 970, 980, or 990
Problem: System error
Solutions:
1.
Take another measurement.
2.
If this is unsuccessful, power down, then power up the unit using the power
button on the right side of the unit. If the error does not recur immediately,
take another measurement.
3.
If the error recurs, remove power from the unit (unplug any power supplies
and remove the battery, if there is one), then reconnect the power. If the
error does not recur immediately, take another measurement.
4.
If the error recurs, contact
SunTech Medical Inc.
or an authorized service
center.
Troubleshooting - Temperature Module
Status Code: 5
Problem: Temperature probe missing or outside of well
Solutions:
1.
Remove and then place the probe in the well.
2.
Ensure that the temperature well (the rear portion of the temperature
module) is well seated on the display. Along the top of the module, the edge
of the well should be flush with the display. Remove and carefully slide the
well onto the display.