80-0067-00-MO-RevF 2017-10-03 SunTech CT40 User Manual | 29
The tested device meets the stated accuracy specification of RMS Error = 1.89% for the range 70% to 100% SaO
2
. The
accuracy is not specifed below 70% SaO
2
. In addition, here are detailed measured A
RMS
values over the specified
discreet ranges:
Range
A
RMS
90-100%
1.54%
80-90%
2.10%
70-80%
2.56%
8. Using SunTech CT40 for Temperature Measurement
The SunTech CT40 has several options for measuring
a patient’s body
temperature.
Covidien
®
FILAC
®
3000 Thermometry Module
Your device may come equipped with a Covidien
®
FILAC
®
3000 digital thermometer capable of taking an oral, axillary
(under arm) or rectal temperature reading. Your thermometer should come with a blue isolation chamber/probe unit
for taking oral and axillary temperature or with a red isolation chamber/probe unit for taking rectal temperature.
NOTE: Always apply a new probe cover before taking temperature. Do not reuse probe cover. Use only probe covers
designed for use with FILAC
®
3000 probes. Using the incorrect probe cover can severely damage the probe and cause
measurement inaccuracies.
NOTE: To change between Celsius and Fahrenheit, press the °C/°F button on the FILAC
®
3000 temperature module.
Press again as needed.