PRELUDE SERVICE MANUAL
PAGE 5
2006
SUNRISE MEDICAL
If the chair is not charging through the charging port, Insure
that the three pin charger socket is properly connected and
in good condition. Then test the voltage coming out of the
joystick (refer to section 1.2) If voltage at joystick charger
socket on the VR2 joystick is less then 18 volts (refer to
section 1.3.) If the voltage is greater then 18 volts refer to
you charger troubleshooting manual.
2.1 Not Charging
A2.1.1
If all of the measurements read full voltage, then replace the controller.
2
3
Section 2
1
Chair is not Charging