QUICKIE S-11 SERVICE MANUAL
PAGE 3
2005
SUNRISE MEDICAL
If the above steps did not correct the problem, change the
main harness.
1.9 Main Harness
1.8 Circuit Breaker Test
To check the circuit breaker set the meter to ohms and
measure the resistance across the circuit breaker as shown
in (figure A1.8.1). If the meter reads more than 1 ohm,
then change the circuit breaker, otherwise proceed to next
step.
A1.8.1
OFF
1000
200
20
2000m
200m
20M
2000K
200K
20K
2000
200
10A
200m
20m
2000u
200u
200
750
V AC
V DC
OHM
A DC
DIGITAL MULTIMETER
V mA COM
10A
If the meter reads more than 1 ohm, then change the circuit breaker.
If the previous steps did not correct the problem, change the main harness.
Check that the battery fuse is in good condition. With the
batteries disconnected remove the fuse cap, inspect the
fuse to see if the fuse is blown. To make sure the fuse is
not blown, set the meter to ohms and measure the resis-
tance across the fuse. see (figure A1.7.1). If the meter
reads more than one ohm, change the fuse, or else pro-
ceed to the next step.
1.7 Battery Fuse
A1.7.1
If the meter reads more than one ohm, change the Battery fuse.
1.6 Battery Wire Harness
Check that the battery wire harness has the polarity correct.
Set the meter to dc volts and measure the connector with
the red lead on the red wire contact (top of the connector)
and the black lead on the black wire contact (bottom of the
connector) as shown in (figure A1.6.1). If both battery wire
harnesses have full voltage and correct polarity, then pro-
ceed to step 1.8. If voltage is absent proceed to step 1.7. If
polarity is reversed correct battery wiring.
A1.6.1
OFF
1000
200
20
2000m
200m
20M
2000K
200K
20K
2000
200
10A
200m
20m
2000u
200u
200
750
V AC
V DC
OHM
A DC
DIGITAL MULTIMETER
V mA COM
10A
If polarity is reversed correct battery wiring.
OFF
1000
200
20
2000m
200m
20M
2000K
200K
20K
2000
200
10A
200m
20m
2000u
200u
200
750
V AC
V DC
OHM
A DC
DIGITAL MULTIMETER
V mA COM
10A