800-433-2797 www.Sun-Med.com
User Guide
Nebutech
®
8660 Reusable Nebulizer
Suggested Cleaning Instructions
for Home Use
The NebuTech Nebulizer is for single-patient use
and should be replaced every 6 months or after
540 cleaning cycles, whichever comes first. Replace
the nebulizer sooner if the it becomes damaged,
contaminated or is not functioning properly.
To Clean the Nebulizer
Follow cleaning instructions after each treatment or
at least once a day.
1. Disassemble all nebulizer components; do not
remove any of the one-way valves.
2. Clean the nebulizer in warm soapy water
(excluding the supply tubing) after each use.
Warm soapy water is defined as a solution of ½
teaspoon of mild liquid detergent for every 1
gallon of water at 35°C (+/- 10°C). Non-scented,
non-lotion, liquid detergent equivalent to Ivory.
3. Submerge the components in the warm soapy
water solution for approximately 60 seconds,
agitating gently.
4. Rinse the devices with similar temperature tap
water.
5. Replace the nebulizer every 6-months, or after
540 cleaning cycles, whichever is first.
WARNINGS/CAUTIONS
1. This product is disposable and is not intended for
prolonged or extended use.
2. All parts should be seated firmly in place.
Improperly assembled nebulizer could prevent
adequate delivery of medication.
3. Intended for single patient, multiple uses.
Do not use on more than one patient.
4. To avoid risk of infection and contamination
follow instruction for use for cleaning
and maintenance of the nebulizer.
5. If the jet is clogged the nebulizer will not produce
adequate mist. This will impact effectiveness of
the treatment. If the nebulizer becomes clogged
and is not functioning properly, replace nebulizer
to ensure adequate medication delivery.
6. Device contains components that may present
choking hazard.
7. If using oxygen, do not use near flame or heat
source.
8. The tubing presents a strangulation hazard.
Always use close supervision when administrating
a treatment to a child. Do not leave a child
unattended during treatment.
9. To reduce the risk of misconnections and patient
injury, always trace tubing from gas source to the
medical device before connecting.
UG-10033 R