54
Follow the instructions on the disinfectant label.
Dry the surface with a non-cotton lint cloth.
7.6 Common Faulty Troubleshooting
Fault Phenomenon
Cause
Solution
Can’t freeze data in
measuring
Lens mode
selection is not
correct
Re-select lens mode
Gain value too high
or too low
Adjust gain value
Can’t reach
automatic
calibration
conditions
Choose manual calibration mode
Can’t collect data
Foot pedal
collection not
connected or
damaged
Re-connected or replace foot pedal, or
use function icon on the screen to
collect
No normal reflection
echogram
Probe not
connected or
damaged
Re-connected or replace probe
For ordinary malfunction users can remove it by themselves. If there are other
malfunctions, please contact our company to solve them.
7.7 Parts Replacement
Customerte can repair or replace the parts information by itself (must use the specified
model of our company)
1. Safety Pipe Type and Rating: F1AH250V (Fuse must meet the requirements of GB
9364)
Note: When replacing the fuse, first disconnect the power supply and open the fuse box
for replacement.
2. A-scan probe (10MHz), Pachymetry probe (20MHz) and B-scan probe (10MHz);
3. Foot switch (model: TFS-1);
Summary of Contents for SK-3000A
Page 12: ...11 1 12 Label This label information is stick on main body of Ophthalmic ultrasound ...
Page 28: ...27 4 5 P scan Test Interface 1 Eye Switch left eye and right eye ...
Page 30: ...29 4 6 B scan Test Interface 1 Eye Switch left eye and right eye ...
Page 35: ...34 ...
Page 41: ...40 2 Click Properties 3 Double click Internet protocol version 4 ICP IPv4 ...