5) Select the B-scan probe type, test mode (B-scan mode / B+A mode), adjust the
gain value (also can adjust during test process)
6) Take down the probe cap, clean the surface of B-scan probe tip with normal
saline,patient lying flat on the examination bed, apply appropriate ultrasound
couplants on eyelid .
7) Step on foot pedal one time or click Start , put the probe on the eyelid vertically , at
the same time , observe the screen and adjust the position of probe until you can
see the retina , lens clearly ,step on the foot pedal again , or click “Stop” to finish a
test.
8) After testing , you can delete bad images and re-test until capture perfect B-scan
ultrasonoscopy.
9) After testing, if you want playback the ultrasonoscopy, just click “Cineloop
sequence” to make it.
10) If you want to do post-processing, please enter the B-scan image
post-processing interface.
5.8 Use Of Calibration Cylinder
1) Turn on main machine power supply.
2) Enter main screen and select “calibrate” mode in lens type drop-down menu .when
calibrate with p scan , set the velocity as 2381m/s.
3) Place calibration cylinder on horizontal table, and drop 1-2 drops of water.
4) Place the probe onto the Calibration Cylinder. The probe should be placed vertically
to the cylinder , step on foot pedal to collect the echogram.
Note
:
Calibration cylinder using environment are within 25°C±5°C interior, the data
measured by cylinder can’t be saved , if you need do next test ,just clear all data.
Summary of Contents for SK-3000A
Page 12: ...11 1 12 Label This label information is stick on main body of Ophthalmic ultrasound ...
Page 28: ...27 4 5 P scan Test Interface 1 Eye Switch left eye and right eye ...
Page 30: ...29 4 6 B scan Test Interface 1 Eye Switch left eye and right eye ...
Page 35: ...34 ...
Page 41: ...40 2 Click Properties 3 Double click Internet protocol version 4 ICP IPv4 ...