MENU ACCESS
7
q
Hold for 3 seconds to access/exit user menu
Accessing user menu
User menu
pH SET POINTS
To modify parameters/maximum pH set point
To LOWER pH - maximum water pH - CONTROLS THE ACID PUMP (optional equipment)
CHLORINE PRODUCTION LEVEL
To modify level
To confirm and change menu
To modify parameters/minumum pH set point
To INCREASE pH - minimum water pH - CONTROLS THE BASE PUMP (optional equipment)
pH PROBE CALIBRATION
Insert probe in pH 7 sample,
wait till reading has stabilized
Clean probe in NEUTRAL.
Insert probe in pH 10
sample and wait till reading
is stable.
REDOX SET POINT
To modify parameters/Redox set point
To confirm and change menu
REDOX PROBE CALIBRATION
Insert probe into 465mv sample,
wait till reading is stable
CONDUCTIVITY SET POINT
To modify parameters/conductivity set point
To confirm and change menu
CONDUCTIVITY PROBE CALIBRATION
Insert the probe in
sample, wait until reading
is stable.
To change menu
and return to Ion menu.
Displays
Conductivity control
5
Redox/Free chlorine control
pH control
Electrolysis level
4
3
2
1
q
Insert the probe in
sample, wait until
reading is stable.
Calibration correct
Repeat calibration or
change probe.
Calibration correct
Repeat calibration
or change probe
RECOMMENDED LEVEL: 6,8
RECOMMENDED LEVEL BETWEEN 7,2 AND 7,5
RECOMMENDED LEVEL: MAXIMUM level according to model (see page2)
DISPLAY
1
DISPLAY
2
DISPLAY
3
DISPLAY
4
1
2
3
4
Calibration correct
Repeat calibration
or change probe