Foot Board Operation Guide
24
CHAPERONE
r
BED EXIT WITH ZONE CONTROL (OPTIONAL EQUIPMENT)
For beds with a scale system:
NOTE:
If the weigh system is in use, it will switch to ”off” when Bed Exit is armed.
1. Before putting the patient on the bed, the weigh system must be zeroed for the Bed Exit System to func-
tion properly (see
for instructions on zeroing the weigh system).
2. Put the patient on the bed and push and release the ”ARM/DISARM” key (top light will come on).
3. The Bed Exit system with Zone Control will automatically select the first zone. To change the zone, push
and hold the “SELECT ZONE” key until the light indicating the desired zone comes on.
4. To deactivate Bed Exit, push the ”ARM/DISARM” key. The selected zone light and ”BED EXIT ON” light will
turn off.
For beds without a scale system:
1. Before putting the patient on the bed, press and hold the ”ARM/DISARM” and the ”SELECT ZONE” keys
together for 5 seconds. The top light will begin to flash.
2. Release the ”ARM/DISARM” and the ”SELECT ZONE” keys and do not touch the bed until the top light
stops flashing.
3. Put the patient on the bed and push and release the ”ARM/DISARM” key (top light will come on).
4. The Bed Exit system with Zone Control will automatically select the first zone. To change the zone, push
and hold the “SELECT ZONE” key until the light indicating the desired zone comes on.
5. To deactivate Bed Exit, push the ”ARM/DISARM” key. The selected zone light and ”BED EXIT ON” light will
turn off.
CHAPERONE
r
ZONE SETTINGS
The first zone (top indicator light) is the traditional Bed Exit zone. The patient can move around the bed freely
but cannot fully exit the bed or the alarm will sound.
The second zone (middle indicator light) is more restrictive than the first zone. When the zone is selected, the
bed measures the location of the patient’s center of gravity. If the patient’s center of gravity moves from the original
location more than 6.5 inches to either side or 13 inches toward the head or foot, an alarm will sound.
The third zone (bottom indicator light) is the most restrictive zone. When the zone is selected, the bed mea-
sures the location of the patient’s center of gravity. If the patient’s center of gravity moves from the original
location more than 1 inch to either side or 1 inch toward the head or foot, an alarm will sound.
NOTE:
All zone dimensions are
±
.5 inches.
WARNING
To avoid possible injury and to assure proper operation when using a powered mattress replacement system
such as XPRT, do not initialize (“arm”) bed exit with Percussion, Vibration, Rotation or Turn
−
Assist active.
The patient motion and position resulting from a dynamic therapy mattress may adversely affect bed exit sys-
tem performance.
Summary of Contents for Secure II 3002
Page 33: ...Optional Pendant Operation 32 ...
Page 46: ...Bed Circuit Boards 45 CPU BOARD 3002 407 950 ...
Page 98: ...Notes 97 ...
Page 101: ...Base Assembly and Options 100 FOOT END HEAD END ...
Page 103: ...Base Assembly and Options 102 HEAD END ...
Page 104: ...Base Assembly and Options 103 HEAD END LEFT SIDE FOOT END ...
Page 105: ...Base Assembly and Options 104 Zoom Option Zoom Option Steer Option ...
Page 106: ...Base Assembly and Options 105 Typical Both Ends Typical Both Ends ...
Page 110: ...109 Lift Assembly Common Assembly part number 3000 200 275 reference only ...
Page 124: ...Notes 121 2 ...
Page 125: ...Base Assembly with Zoomr Option 122 HEAD END ...
Page 127: ...Base Assembly with Zoomr Option 124 HEAD END HEAD END ...
Page 128: ...Base Assembly with Zoomr Option 125 FOOT END 8 Caster Option 6 Caster Option ...
Page 129: ...Base Assembly with Zoomr Option 126 FOOT END HEAD END ...
Page 130: ...Base Assembly with Zoomr Option 127 ...
Page 131: ...Base Assembly with Zoomr Option 128 ...
Page 139: ...Base Assembly with Battery Back Up Option 136 HEAD END HEAD END 2025 31 805 Ground Strap ...
Page 140: ...Base Assembly with Battery Back Up Option 137 ...
Page 141: ...Base Assembly with Battery Back Up Option 138 ...
Page 144: ...Litter Assembly 141 ...
Page 145: ...Litter Assembly 142 ...
Page 146: ...Litter Assembly 143 ...
Page 147: ...Litter Assembly 144 DETAIL A Standard Length Bed 8 Shorter Length Bed ...
Page 149: ...Litter Assembly 146 DETAIL F ...
Page 153: ...Electrical Litter Assembly 150 ...
Page 154: ...Electrical Litter Assembly 151 Siderail Cable Attachment ...
Page 155: ...Electrical Litter Assembly 152 DETAIL C To Gatch Motor See Detail C ...
Page 156: ...Electrical Litter Assembly 153 DETAIL A DETAIL B 115V BED 230V BED ...
Page 157: ...Electrical Litter Assembly 154 SMART TV OPTION UNITED KINGDOM POWER CORD CABLE INSTALLATION ...
Page 159: ...Mechanical Litter Assembly 156 ...
Page 160: ...Mechanical Litter Assembly 157 ...
Page 162: ...Mechanical Litter Assembly 159 SECTION A A A A ...
Page 164: ...Notes 161 ...
Page 169: ...166 Fowler Drive Assembly Standard Fowler Drive Shown ...
Page 178: ...Notes 175 ...
Page 180: ...Optional Scale Bed Exit Assembly 177 FOOT END ...
Page 182: ...Optional HWI Assembly with Communication 179 ...
Page 183: ...Optional HWI Assembly w Communication Ports 180 ...
Page 184: ...Optional HWI Assembly w Communication Ports 181 ...
Page 187: ...Zoom Option Litter Assembly 184 DETAIL A Assembly part number 3002 131 10 reference only ...
Page 188: ...Zoom Option Litter Assembly 185 DISPLAY HOUSING ...
Page 189: ...Zoom Option Litter Assembly 186 On Off Switch Inside Enclosure ...
Page 190: ...Zoom Option Litter Assembly 187 ...
Page 191: ...Zoom Option Litter Assembly 188 ...
Page 194: ...Notes 191 ...
Page 198: ...Battery Back Up Option Head End Enclosure 195 ...
Page 200: ...Head End Siderail Assembly 197 LEFT SIDERAIL SHOWN ...
Page 201: ...Head End Siderail Assembly 198 CARRIER ARM ASSEMBLY LEFT SIDERAIL SHOWN ...
Page 203: ...Head End Siderail Assembly 200 SIDERAIL INNER PANEL ASSEMBLY ...
Page 204: ...Head End Siderail Assembly 201 LEFT HEAD END SIDERAIL WIRE ROUTING ...
Page 205: ...Head End Siderail Assembly 202 RIGHT HEAD END SIDERAIL WIRE ROUTING ...
Page 257: ...2040 120 4 Optional Pleur Evac Rack w Defib Tray 252 ...
Page 263: ...258 3001 333 000 Optional Accessory Adapter Frame ...
Page 268: ...3001 318 20 Optional Bed Extender Assembly 263 C F H C D C D E J ...