8. Using diagonal pliers, cut the cable tie (3000-300-114) (AK) that secures the 37-pin cable (M) to the cable routing tray
(D) (Figure 9). Discard the cable tie.
F
Fiig
gu
urre
e 9
9 –– R
Re
ep
plla
acciin
ng
g tth
he
e b
ba
atttte
erryy ((lle
efftt))
9. Move the 37-pin cable (M) and cable routing tray (D) up to remove the battery (700000341245) (V) from the body (C).
10. Unplug the two battery cables and discard the battery.
N
No
otte
e -- When you reinstall, make sure that the red wire connects to positive (+) and the black wire connects to negative
(-) on the battery.
11. Reverse steps to reinstall.
12. Verify proper operation before you return the product to service.
EN
18
5212-009-201 Rev A.0