No nurse call communication cable connection (37 pin
cable is not detected)
1. Check the nurse call communication cable connection
at the S
Se
eccu
urre
e C
Co
on
nn
ne
ecctt and at the wall connection.
2. Check the nurse call communication cable for damage.
No connection to the product
1. Power cycle the product and the S
Se
eccu
urre
e C
Co
on
nn
ne
ecctt.
2. Check that the brakes are applied.
Low battery alert
1. Check that the power supply is plugged into the S
Se
eccu
urre
e
C
Co
on
nn
ne
ecctt and to the wall outlet.
2. Check the power supply for damage.
3. Check the DC output voltage from the power supply.
4. Replace the battery. See
(page 17) or
(37-pin connector, right side)
Nurse call actions or alarms are not working
1. Power cycle the product and the S
Se
eccu
urre
e C
Co
on
nn
ne
ecctt.
2. Check that the brakes are applied.
3. Check that the headwall and S
Se
eccu
urre
e C
Co
on
nn
ne
ecctt dip-
switches are set for the nurse call system and options.
4. Check that the headwall settings are configured for the
nurse call system in the product maintenance
configuration menu.
EN
16
5212-009-201 Rev A.0