21
4. On the Advanced Imaging Modality screen, press the
IRIS mode button
.
4
Note:
When the IRIS mode button is red, the L10 AIM Light Source
is properly connected to the IRIS Ureteral Kit and IRIS mode can be
activated. If the IRIS mode button is gray, required equipment may not
be properly connected and IRIS mode is not available. Refer to the Setup
section.
5. On the IRIS mode screen, press the
touchscreen power button
so it
appears green, which indicates the mode is activated.
Note:
Press the touchscreen power button again to disable IRIS mode;
the button will appear gray.
1
7
6
8
5
6. Adjust the
IRIS brightness level
by pressing the plus or minus buttons.
7. Press the solid line or the dotted line to toggle the IRIS light output
between the
continuous or pulsating
setting, respectively.
•
In the continuous setting, the laser energy is continuous.
•
In the pulsating setting, the laser energy pulses in a repeating pattern
of 0.5 second on/0.5 second off (1 pulse cycle per second).
8. Press the
Home button
to return to the White Light screen
.