Operation
Relocating the support surface
Relocating the support surface is for the purpose of cleaning or moving to another InTouch bed frame.
WARNING
•
Do not move the support surface without assistance.
•
Do not use the support surface as a transfer device to avoid the risk of patient injury.
Prerequisite:
Minimum of two operators are required to relocate the support surface and setup this product.
1.
Make sure that there is not a patient on the support surface.
2. Remove the linens.
3. Unplug the power cord for the support surface from the foot end auxiliary power outlet.
4. Unplug the support surface connector.
5. Remove the headboard and footboard
6. Position an operator at each end of the support surface and remove the support surface and relocate.
Note:
Use all four handles when relocating the support surface.
Note:
See (
) of the support surface.
7.
Reinstall the
InTouch
headboard and footboard.
Transferring of patient from one patient support platform to another
WARNING
•
Do not use the support surface as a transfer device to avoid the risk of patient injury.
•
Do not stick needles into a support surface through the support surface cover. Holes may allow body fluids to enter
the inside (inner core) of the support surface and could cause cross-contamination, product damage, or product
malfunction.
•
Do not exceed the safe working load of the
InTouch
bed frame when supporting both the patient and the support
surface. Excess weight could cause unpredictable safety and performance of this product.
To transfer the patient from one patient support surface to another:
Prerequisite:
Follow hospital protocols required to transfer a patient from one surface to another.
1.
Position one patient support platform alongside the other patient support platform while minimizing the gap between
the two platforms.
2. Set the brakes to on for both patient support platforms.
3. Activate maximum inflation of the support surface until the inflation level is reached. See the
InTouch
Starting and
Stopping Max Inflate section of the manual for use instructions.
4. Adjust the patient support platform heights so that they are level with one another.
5. Transfer the patient following all applicable safety rules and institution protocols for patient and operator safety.
6. Turn off Max Inflate.
Summary of Contents for Isolibrium 2971
Page 2: ...sample text ...
Page 6: ...sample text ...
Page 80: ...E EN N 1 72 2971 109 001 REV B www stryker com Isolibrium assembly ...
Page 81: ...E EN N www stryker com 2971 109 0 01 REV B 1 73 Isolibrium assembly ...
Page 82: ...E EN N 1 74 2971 109 001 REV B www stryker com Isolibrium assembly ...
Page 86: ...E EN N 1 78 2971 109 001 REV B www stryker com Footbox assembly Torque item A to 22 4 in lb ...
Page 87: ...E EN N www stryker com 2971 109 0 01 REV B 1 79 Footbox assembly Torque item H to 22 4 in lb ...
Page 89: ...E EN N www stryker com 2971 109 0 01 REV B 1 81 Footbox assembly Torque item C to 11 2 in lb ...
Page 90: ...E EN N 1 82 2971 109 001 REV B www stryker com Footbox assembly Torque item B to 22 4 in lb ...
Page 104: ...sample text ...
Page 106: ...sample text ...
Page 177: ...F Fr ra an nç ça ai is s F FR R www stryker com 2971 109 0 01 REV B 2 69 2971 021 001 ...
Page 178: ...F Fr ra an nç ça ai is s F FR R 2 70 2971 109 001 REV B www stryker com 2971 021 001 ...
Page 179: ...F Fr ra an nç ça ai is s F FR R www stryker com 2971 109 0 01 REV B 2 71 2971 021 001 ...
Page 180: ...F Fr ra an nç ça ai is s F FR R 2 72 2971 109 001 REV B www stryker com 2971 021 001 ...
Page 201: ...sample text ...