Installation
The product must reach the ambient operating temperature before installation or operation to avoid the risk of permanent
damage to this product. Upon initial installation, reset the CPR.
WARNING
•
Risk of Shock. This product must only be connected to a supply mains with protective earth. Grounding reliability can
be achieved only when a hospital grade receptacle is used.
•
Explosion risk. This product is not suitable for use in the presence of flammable anesthetic mixture with air or with
oxygen or nitrous oxide other than nasal, mask type, or half bed-length tent type.
•
Do not move the support surface without assistance.
•
Do not operate the support surface without the
Isolibrium
top and bottom covers and linens.
•
Always position the support surface on the
InTouch
bed frame to that the support surface does not contact the
headboard.
•
Do not use when gaps between the support surface and the bed frame are present to prevent the risk of entrapment.
Only use the
Isolibrium
support surface with
InTouch
bed frames.
•
Do not use fitted sheets with this support surface.
•
Do not attach the power cord to any moving parts of the
InTouch
bed frame. Improper routing of the power cord
could cause the risk of electric shock, pinching the power cord, or interfere with bed functions.
•
Always confirm proper functionality of the scale system after you install the support surface to prevent product
malfunction.
•
Do not use extension cords with
Isolibrium
. It is only intended to be powered by
InTouch
with the power cord
supplied.
•
Always use the power source provided by the
InTouch
foot end auxiliary power outlet with
Isolibrium
.
Note:
Do not plug into the power source until the assembly is complete.
Prerequisite
: Minimum of two operators are required to install the support surface.
1.
Put the
InTouch
bed into the flat position before installing the support surface.
2.
If powered, unplug the
InTouch
from the wall outlet and turn the battery switch to off.
3.
Remove the
InTouch
headboard and footboard from the bed.
4.
Position an operator at each end of the support surface and lift onto the
InTouch
bed frame.
5.
Reinstall the
InTouch
headboard and footboard.
6.
Position the support surface against the footboard. Allow a small air gap at the headboard.
7.
Connect the integration cable from the support surface foot end to the
InTouch
cable adapter (B) (
8.
Turn the locking collars clockwise to secure the connection.
9.
Place the linens on the support surface as loose as possible to avoid creating a
hammocking
effect and increasing
interface pressures.
10. Plug the power cord for the support surface into the
InTouch
foot end auxiliary power outlet (A) (
Summary of Contents for Isolibrium 2971
Page 2: ...sample text ...
Page 6: ...sample text ...
Page 80: ...E EN N 1 72 2971 109 001 REV B www stryker com Isolibrium assembly ...
Page 81: ...E EN N www stryker com 2971 109 0 01 REV B 1 73 Isolibrium assembly ...
Page 82: ...E EN N 1 74 2971 109 001 REV B www stryker com Isolibrium assembly ...
Page 86: ...E EN N 1 78 2971 109 001 REV B www stryker com Footbox assembly Torque item A to 22 4 in lb ...
Page 87: ...E EN N www stryker com 2971 109 0 01 REV B 1 79 Footbox assembly Torque item H to 22 4 in lb ...
Page 89: ...E EN N www stryker com 2971 109 0 01 REV B 1 81 Footbox assembly Torque item C to 11 2 in lb ...
Page 90: ...E EN N 1 82 2971 109 001 REV B www stryker com Footbox assembly Torque item B to 22 4 in lb ...
Page 104: ...sample text ...
Page 106: ...sample text ...
Page 177: ...F Fr ra an nç ça ai is s F FR R www stryker com 2971 109 0 01 REV B 2 69 2971 021 001 ...
Page 178: ...F Fr ra an nç ça ai is s F FR R 2 70 2971 109 001 REV B www stryker com 2971 021 001 ...
Page 179: ...F Fr ra an nç ça ai is s F FR R www stryker com 2971 109 0 01 REV B 2 71 2971 021 001 ...
Page 180: ...F Fr ra an nç ça ai is s F FR R 2 72 2971 109 001 REV B www stryker com 2971 021 001 ...
Page 201: ...sample text ...