Suggested disinfectants:
•
Quaternaries (active ingredient - ammonium chloride) that contain less than 3% glycol ether
•
Phenolic solution (Matar)
•
Chlorinated bleach solution (6500 ppm)
•
Potassium peroxymonosulfate
•
Accelerated hydrogen peroxide
•
70% isopropyl alcohol
To disinfect the support surface covers after each patient use, follow these steps in order:
1. Inflate the support surface with the pump.
N
No
otta
a -- Follow steps 1 and 2 when you use the
2. Disconnect the mains power from the receptacle.
N
No
otta
a -- Leave the hose connected to the support surface to retain the air.
3. Clean and dry the support surface covers before disinfectants are applied.
4. Apply recommended disinfectant solution with a damp cloth or pre-soaked wipes. Do not soak the support surface.
N
No
otta
a -- Make sure that you follow the disinfectant manufacturer’s instructions for appropriate contact time and rinsing
requirements.
5. Rinse the product with a clean, damp cloth to remove excess disinfectant.
6. Dry the product with a clean, dry cloth to remove any excess liquid or disinfectant.
7. Allow the product to dry completely before you return to service.
T
Trro
ou
ub
blle
essh
ho
oo
ottiin
ng
g
A
AT
TT
TE
EN
NZ
ZIIO
ON
NE
E -- Do not modify the product or any components of the product. Modifying the product can cause
unpredictable operation resulting in injury to patient or operator. Modifying the product also voids its warranty.
Service only by qualified personnel.
P
Prro
ob
blle
em
m
C
Ca
au
usse
e
R
Re
ecco
om
mm
me
en
nd
de
ed
d a
accttiio
on
n
Pressure alarm indicator blinking on
the 2940 IIsso
oA
Aiirr pump
1. Bad CPR connection
2. Bad hose connection at a fitting
3. A hole in an air cell or air hose
1. Inspect the CPR connection
1a. Listen for leaks around the CPR
connection
2. and 3. Inspect all air hoses, fittings,
and air cells
N
No
otta
a -- The A and B air circuits on the 2940 IIsso
oA
Aiirr pump are opposite from the A and B air cells on the 2941 IIsso
oA
Aiirr support
surface (Figura 2).
2941-009-020 Rev E.0
7