Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The Epic II Critical Care Bed is suitable for use in the electromagnetic environment specified below. The customer
or the user of the Epic II Critical Care Bed should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic
Environment Guidance
Electrostatic Discharge (ESD)
IEC 61000-4-2
+6 kV contact
+8 kV air
+6 kV contact
+8 kV air
Floors should be wood,
concrete, or ceramic tile.
If floors are covered with
synthetic material, the relative
humidity should be at least
30%.
Electrostatic fast
Transient/burst
IEC61000-4-4
+2 kV for power
supply lines
+1 kV for input/
output lines
+2 kV for power
supply lines
+1 kV for input/
output lines
Main power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
+8 kV differential mode
+2 kV common mode
+8 kV differential mode
+2 kV common mode
Main power quality is that of
a typical commercial and/or
hospital environment.
Voltage dips, voltage
variations and short
interruptions on power supply
input lines
IEC 61000-4-11
<5%Ut (95% dip in Ut) for
0,5 cycle
40%Ut (60% dip in Ut) for
5 cycles
70%Ut (30% dip in Ut) for
25 cycles.
<5% Ut (>95% dip in Ut)
for 5 sec.
<5%Ut (95% dip in Ut)
for 0,5 cycle
40%Ut (60% dip in Ut)
for 5 cycles
70%Ut (30% dip in Ut)
for 25 cycles.
<5% Ut (>95% dip in Ut)
for 5 sec.
Main power quality should be
that of a typical commercial
and/or hospital environment. If
the user of the Epic II Critical
Care Bed
requires continued
operation during power main
interruptions, it is recommended
that the device be powered
from an uninterrupted power
supply or a battery.
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
and/or hospital environment.
Note: U
T
is the a.c. mains voltage prior to applications of the test level.